Long-Term Neural Spiking and Decoding Performance in BrainGate Clinical Trials
A study published in Nature documents the performance of intracortical microelectrode arrays in people with implanted brain–computer interfaces over a 20-year period, revealing that neural spiking activity persists for multiple years with little average decline.
The research, based on the BrainGate and BrainGate2 clinical trials conducted between 2004 and 2019, involved 14 participants, each with one or two 96-channel platinum-metalized Utah microelectrode arrays implanted in their cortex. Of the 20 total arrays, 19 were placed in the hand area of the motor cortex, while one was placed in the middle frontal gyrus.
The trials were designed to evaluate the safety of the implanted microelectrode arrays and to uncover fundamental insights into the effective design of intracortical BCIs. Testing new neural decoding architectures and task designs was common throughout the trials, making it impractical to summarize all data types collected.
To assess performance, researchers curated two datasets: a “yield dataset” containing a single neural recording from every research session to estimate array yield, and a “decoding dataset” consisting of neural recordings during which participants used a BCI to move a computer cursor toward defined targets. The yield dataset included 2,289 recording sessions, while the decoding dataset had 729 sessions. Total array implant days ranged from 296 to 2,780, with a median of 774 days.
Five participants—S1, S2, S3, A1, and T1—were excluded from decoding analyses because high-bandwidth neural data (30 kHz) required for uniform evaluation was inconsistently available. Prior publications had demonstrated successful neural cursor control in these participants, but the current study focused on standardized data.
The array in the middle frontal gyrus of participant T10 was used only for recording evaluation, as its neural tuning properties were not comparable to the hand knob area.
The study’s findings highlight the long-term viability of these implants, with neural activity remaining stable over extended periods. This stability is critical for the development of reliable BCIs, which aim to restore communication and mobility for people with severe paralysis.